• Αθ. Σταγειρίτη 7, Πυλαία Θεσ/νίκης
  • +30 2310 942346

Ποιότητα

σε υλικά & κατασκευή

Σωληνωτοί Κλίβανοι

High Temperature Tube Furnace, up to 1500˚C - Universal Mounting - SiC Heaters

Κατηγορίες: Κωδικός:
RCT-AS1-T-1500 and RCT-AS1-O-1500 tube furnace families were designed to offer a low cost alternative as compared with our MoSi2 heaters furnaces for applications requiring heat treatment up to 1500˚C continuously.   
Utilizing the unique high density Kanthal SiC resistors these furnaces have superior performance in terms or resistance to oxidation highly extending the workable life of your investment. All SiC heating resistors will increase in resistance over time at elevated temperatures. Due to the unique nature of the high density Kanthal SiC resistors this effect is severally suppressed.
Offered only as a single zone model will give repeatable and reliable results with many years of workable life operating horizontally on its feet or vertically using an appropriate stand.
Equipped as a standard with our advanced remote data acquisition and control system PYROMODULAR and with the addition of an appropriate reaction tube these tube furnaces can be used for many applications requiring ultra high temperature.
  • Maximum Continuous Temperature 1500 ˚C.
  • Hinged split tube or closed tunnel construction.  
  • Kanthal SiC, exposed type resistors.
  • Highly resistant in thermal shock.
  • Universal, vertical and horizontal operation.
  • Excellent price per performance ratio – Low cost.

Βασικά χαρακτηριστικά

  • Best available quality KANTHAL helical cut SiC resistors driven by THERMANSYS PCC control platform insure furnace long life operation up to 1500˚C continuous operation. 
  • Low mass vacuum formed thermal insulation enables high output available for the load and fast heat up rates while significantly contributes to energy savings under daily thermal cycling.
  • Control strategy focusing in high power factor for all workable temperature areas insures compliance with EMC (Electro-Magnetic Compatibility) standards.
  • Vertical and horizontal mounting flexibility covers different present and future laboratory needs.
  • Accurate and uniform temperature profiles.
  • Modern double wall construction keeps external surfaces temperature low, emphasizing in operator safety. Internal skin is exclusively made from stainless steel to enhance durability.
  • Ergonomic design with no protruding edges, bolts or other features combines stainless steel parts with painted finish parts for an improved esthetic result.
  • Temperature PID control.
  • USB port for connection to a PC
  • Touch Screen Computer running PYROLOGISM software human-machine interface.
  • 3 user process thermocouple inputs available (B, E, J, K, N, R, S, T type- software configurable).
  • Power and true RMS Current measuring circuits.
  • Heater failure, open control thermocouple detection, door open alarms and interlocks. Alarms and events front panel led array.
  • Stand alone over-temperature limiter (Watchdog) with manual reset in accordance with EN 60519-2 to protect the oven and load. Overrides main controller and cut off heater power if user adjustable high limit is reached.

Προδιαγραφές

Standard features:
  • Maximum continuous temperature 1500 ˚C.
  • Operating Power: 208 /240VAC – 50/60Hz.
  • R type embedded thermocouples.
  • Mounting orientation: Horizontal, and vertical.
  • Temperature control accuracy ±1 ˚C.
  • Exposed resistors type
  • Single heating zone configuration models.
  • Each zone is equipped with two independed thermocouples for controller and over-temperature limiter feedback.
  • Thermocouple inputs:3 chan. - B, E, J, K, N, R, S, T type -software configurable, 24 bit A/D conversion, 0-45°C cold junction compensated, Typical accuracy ±0.2% f.s @ 25 , resolution 0.1 °C.
Optional features:
  • Remote, touch screen temperature computer, running the specially designed PYROLOGISM 2.0 software on a 10.0in Tablet PC
Add suffix _TSC
  • Programmable stand-alone over-temperature limiter (Watchdog) with manual reset in accordance with EN 60519-2 to protect the heater and load,
Add suffix _WD
  • UPS (Uninterrupted Power Supply) that will keep system alive for short periods of power failure and restore program after power recovery
Add suffix _UPS


CE Certified. Compliant with Low Voltage Directive 2006/95/EC (harmonized referenced standard EN 61010-1: 2001 and EN 61010-2-010:2003) and EMC Directive 2004/108/EC (harmonized referenced standard EN 61326-1:2006).

Produced in GREECE following ISO 9001:2008 quality management system and ISO 14001:2004 environmental management system.

Μοντέλα

For detailed ordering and technical information please see our technical brochures:
Closed Tube, Maximum Temperature 1500˚C, Universal Operation.
Model No Max. Temperature (C) Heated Diam. x Length
RCT-AS1-T-D6L20-1500 1500 60x200mm
RCT-AS1-T-D6L40-1500 1500 60x400mm
RCT-AS1-T-D6L60-1500 1500 60x600mm
RCT-AS1-T-D9L40-1500 1500 90x400mm
RCT-AS1-T-D9L60-1500 1500 90x600mm

Doc No. RCT-AS1-T-1500 November 2016
Split Tube, Maximum Temperature 1500˚C, Universal Operation.
Model No Max. Temperature (C) Heated Diam. x Length
RCT-AS1-O-D5L20-1500 1500 50x200mm
RCT-AS1-O-D5L30-1500 1500 50x300mm
RCT-AS1-O-D5L60-1500 1500 50x600mm
RCT-AS1-O-D8L40-1500 1500 80x400mm
RCT-AS1-O-D8L60-1500 1500 80x600mm

Doc No. RCT-AS1-O-1500 January 2019